Online Conductivity Controller LOCC-A10 features 3 points conductivity calibration system, TDS measurement mode, and 4 to 20 mA output signal transmitter. The large and clear back light LCD displays Calibration and setup values, Temperature values, ATC (automatic temperature compensation) values, high and low limit of measurement. This controller measures Conductivity and TDS values in µS/ cm, and mS/ cm units.
| Conductivity measurement range | 0.01 to 20.00, 200.0, 2000 µS/ cm, 20.00, 200.0 mS/ cm |
| Conductivity calibration points | 1 to 3 points |
| Conductivity accuracy | ±1 % F.S. |
| Conductivity resolution | 0.001/ 0.01/ 0.1/ 1 |
| Conductivity calibration solutions | 84 µS/ cm, 1413 µS/ cm, 12.88 mS/ cm,111.8 mS/ cm |
| TDS measurement range | 0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt) |
| TDS accuracy | ±1 % F.S. |
| TDS factor | 0.1 to 1.0 |
| Temperature measurement range | 0 to 100°C, 32 to 212°F |
| Temperature accuracy | 1°C |
| Temperature resolution | 0.1°C |
| Temperature calibration points | 1 point |
| Temperature calibration range | Measured value ± 10°C |
| Signal output | 4 to 200 mA |
| Load | Max. 500 Ω |
| Low and high alarm limits | 0.02 uS/ cm to 20.0 mS/ cm, selectable |
| Interface | RS485 |
| Connection terminals | Detachable screw terminals |
| Temperature compensation | Automatic, 0 to 100°C , 32 to 212°F |
| Cell constant | K= 0.1, 1, 10 |
| Environmental conditions | Ambient temperature to ≤ 60°C, RH ≤ 80% |
| Reference temperature | 25°C |
| Display | LCD |
| Power | DC 24 V |
| Dimension (L*W*H) | 96 × 96 × 75 mm |
| Weight | 350 g |
Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.
Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.
Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.
Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.
Popular Categories
Navigation
Contact Us Map