Ion chromatograph system LICS-A22 with cation and anion dual channel design, independently operates and simultaneously detects both channels without disturbing each other. Temperature controlled bipolar conductivity detector, improves the detection performance and stability of the instrument. System is comprised of an auto sampler, anion and cation suppressor.

Ion Chromatographic pump <table class='table table-striped table-bordered table-responsive'><tr><td><strong>Type</strong></td><td>High pressure and low pulse double plunger tandem advection pump</td></tr><tr><td><strong>Maximum pressure</strong></td><td>35 MPa (PEEK)</td></tr><tr><td><strong>Display pressure accuracy</strong></td><td>≤ 0.1 MPa</td></tr><tr><td><strong>Flow stability</strong></td><td>(0.2 to 0.5) ml/ min ≤ 3 % , (0.5 to 1.0) ml/ min ≤ 2 %, > 1.0 ml/ min ≤ 2 %</td></tr><tr><td><strong>Flow range</strong></td><td>0.001 to 9.999 ml/ min</td></tr><tr><td><strong>Pressure pulse</strong></td><td>≤ 0.5 %</td></tr></table>
Numerical control and electromagnetic injection <table class='table table-striped table-bordered table-responsive'><tr><td><strong>Maximum pressure</strong></td><td>35 MPa</td></tr><tr><td><strong>Contact material of the rotor</strong></td><td>PEEK</td></tr><tr><td><strong>Control mode</strong></td><td>By Stepper motor</td></tr><tr><td><strong>Power supply</strong></td><td>DC 24 V</td></tr></table>
Column heater <table class='table table-striped table-bordered table-responsive'><tr><td><strong>Operating temperature range</strong></td><td>Room temperature 20 to 60°C ( 68 to 140°F)</td></tr><tr><td><strong>Controlling temperature accuracy</strong></td><td>± 0.01°C</td></tr><tr><td><strong>Allowable deviation of column heater temperature</strong></td><td>± 2°C</td></tr><tr><td><strong>Temperature stability</strong></td><td>≤ 1°C/ h</td></tr></table>
Flow system <table class='table table-striped table-bordered table-responsive'><tr><td><strong>Plastic flow path</strong></td><td>Made up of PEEK material</td></tr><tr><td><strong>Six way valve</strong></td><td>PEEK material pressure 5000 psi, independent automatic collection and flow function</td></tr></table>
Thermal buffer system for Eluent <table class='table table-striped table-bordered table-responsive'><tr><td><strong>Temperature range</strong></td><td>25 to 40°C ( 77 to 104°F)</td></tr></table>
Degassing device <table class='table table-striped table-bordered table-responsive'><tr><td><strong>Type</strong></td><td>Built-in low pressure</td></tr><tr><td><strong>Vacuum degree</strong></td><td>(-) 70 kPa</td></tr><tr><td><strong>Maximum flow rate</strong></td><td>10 ml/ min</td></tr><tr><td><strong>Internal volume</strong></td><td>30 µl</td></tr><tr><td><strong>Degassing efficiency</strong></td><td>10 ml/ min 90 %</td></tr></table>
Digital temperature and control detection system <table class='table table-striped table-bordered table-responsive'><tr><td><strong>Type</strong></td><td>Temperature-control and Bipolar conductivity detector</td></tr><tr><td><strong>Detection mode</strong></td><td>Bipolar conductivity detection</td></tr><tr><td><strong>Cell volume</strong></td><td>≤ 0.8 µl</td></tr><tr><td><strong>Detection range</strong></td><td>0 to 50000 µS/ cm</td></tr><tr><td><strong>Output voltage</strong></td><td>-6000 to +6000 mV (adjustable)</td></tr><tr><td><strong>Operating temperature</strong></td><td>Room temperature 5 to 60°C ( 41 to 140°F)</td></tr><tr><td><strong>Controlling temperature accuracy</strong></td><td>± 0.01°C</td></tr><tr><td><strong>Maximum pressure</strong></td><td>10 MPa</td></tr><tr><td><strong>Linear range</strong></td><td>≥ 103</td></tr><tr><td><strong>Instrument linearity</strong></td><td>≥ 0.999</td></tr><tr><td><strong>Quantitative repeatability</strong></td><td>≤ 1.0 %</td></tr><tr><td><strong>Qualitative repeatability</strong></td><td>≤ 1.0 %</td></tr></table>
Eluent generator <table class='table table-striped table-bordered table-responsive'><tr><td><strong>Eluent type</strong></td><td>KOH/ NaOH</td></tr><tr><td><strong>Eluent concentration range</strong></td><td>0.1 to 100 mM</td></tr><tr><td><strong>Concentration increment</strong></td><td>0.1 mM</td></tr><tr><td><strong>Flow rate change</strong></td><td>0.1 to 3.0 ml/ min</td></tr><tr><td><strong>Maximum pressure</strong></td><td>20 MPa</td></tr><tr><td><strong>Minimum pressure</strong></td><td>5 MPa</td></tr></table>
Power 150 W
Power supply DC 220 V, 60 Hz
External dimension 500 × 500 × 760 mm
Net weight 48 kg
Gross weight 73 kg
HS code 9027201200
  • 35 MPa maximum pressure ion chromatographic Pump (PEEK material)
  • Auto sampler with (2 ml × 120 samples- Injection unit)
  • Temperature-controlled bipolar conductivity detector
  • Built-in low pressure degassing technology
  • Advanced thermal buffer system for eluent
  • Intelligent (PEEK) flow path system with automatic cleaning mode
  • One key operation to complete flow path switch
  • Highly stable and efficient performance
Used to detect small molecules (in the form of anion, cation) in water testing & processing plants, food, chemical and dye industry, mining, metallurgy, and other fields.

Ion Chromatography

Ion chromatograph system LICS-A21

Ion chromatograph system LICS-A21

  • Ion Chromatographic Pump
  • TypeHigh pressure and low pulse double plunger tandem advection pumpMaximum pressure35 MPa (PEEK)Pressure display accuracy≤ 0.1 MPaFlow stability(0.2 to 0.5) ml/ min ≤ 3 % (0.5 to 1.0) ml/ min ≤ 2 % > 1.0 ml/ min 2 %Flow range0.001 to 9.999 ml/ minPressure pulse≤ 0.5%
  • Conductivity detector
  • TypeTemperature Control and Bipolar Conductivity DetectorDetection modeBipolar conductivity detectionCell volume≤ 0.8 µLDetection range0 to 45000 µS/ cmDetection resolution≤ 0.0020 nS/ cmOutput voltage- 6000 to +6000 mV (adjustable)Baseline noise≤ 0.05 % F.S.Baseline drift≤ 3 % FS/ 30 minOperating temperature rangeRoom temperature +5 to 60°C (41 to 140°F)Controlling temperature accuracy± 0.01°CMaximum pressure10 MPaLinear range≥ 103Instruments linearity≥ 0.999Quantitative repeatability≤ 1.0 %Qualitative repeatability≤ 1.0 %Minimum detectable concentrationCl ≤ 0.0005 ug/ ml BrO3 ≤ 0.001 ug/ ml Li+ ≤ 0.001 ug/ ml
  • Flow system
  • Plastic flow pathMade of PEEK materialsSix way ValvePEEK material, 5000 psi pressure, independent automatic collection and flow function
Ion chromatograph system LICS-A20

Ion chromatograph system LICS-A20

  • Ion Chromatographic Pump
  • TypeHigh pressure and low pulse double plunger tandem advection pumpMaximum pressure35 MPa (PEEK)Pressure display accuracy≤ 0.1 MPaFlow stability(0.2 to 0.5) ml/ min ≤ 3 % (0.5 to 1.0) ml/ min ≤ 2 % > 1.0 ml/ min 2 %Flow range0.001 to 9.999 ml/ minPressure pulse≤ 0.5%
  • Conductivity detector
  • TypeTemperature Control and Bipolar Conductivity DetectorDetection modeBipolar conductivity detectionCell volume≤ 0.8 µLDetection range0 to 35000 µS/ cmDetection resolution≤ 0.0020 nS/ cmOutput voltage- 6000 to +6000 mV (adjustable)Baseline noise≤ 0.05 % F.S.Baseline drift≤ 3 % FS/ 30 minOperating temperature rangeRoom temperature +5 to 60°C (41 to 140°F)Controlling temperature accuracy± 0.01°CMaximum pressure10 MPaLinear range≥ 103Instruments linearity≥ 0.999Quantitative repeatability≤ 1.0 %Qualitative repeatability≤ 1.0 %Minimum detectable concentrationCl ≤ 0.0005 ug/ ml BrO3 ≤ 0.001 ug/ ml Li+ ≤ 0.001 ug/ ml
  • Flow system
  • Plastic flow pathMade of PEEK materialsSix way ValvePEEK material, 5000 psi pressure, independent automatic collection and flow function

Gas Chromatography

Gas Chromatography LGC-B11

Gas Chromatography LGC-B11

  • Oven
  • Maximum capacity22 LTemperature rangeRT + 5 ℃ to 400 ℃Temperature accuracy± 0.1 ℃Temperature program9-phase/ 10 platformProgram total time9999.9 minMaximum temperature heating rate0.1 to 60 ℃/ minMaximum cooling rate˂ 10 mins (250 to 50 ℃)
  • Sample injector
  • Temperature rangeRT+ 7 ℃ to 420 ℃Temperature controlAutomaticInjector typeColumn or splitMaximum sample inlet3 pcsCarrier gas flow typeConstant pressurePre column pressure range0 to 400 kPaPre column pressure accuracy0.1 kPaMobile phase flow rangeH2: 0 to 200 ml/min, N2: 0 to 150 ml/min
  • Detector
  • Temperature rangeRT + 7 ℃ to 420 ℃Maximum installationDual detectorFlame ignition modeAutomaticFlame Ionization Detector (FID)Logarithmic amplifier High voltage switch control Baseline signal display Ignition coil control Detection limit: ≤3 ×10 to 12 g/s (Sample: n-hexadecane) Baseline noise: ≤ 5 × 10 - 14 A Baseline draft: ≤ 6 × 10 -13 A Dynamic range: 10 7 RSD: ≤ 3 %

GC-MS

Gas chromatography mass spectrometry GC-MS-879

Gas chromatography mass spectrometry GC-MS-879

  • Gas Chromatography Specifications
  • Model NoGC-87Flow rate1 ~ 1.5 ml / minMaximum flow rate10 ml / minInlet temperature450 °CPressure range0 ~ 100 psiHeating rateUpto 120 °C / minRoom temperature4 °C ~ 450 °CPressure control modeElectronic pressure control ( Supports CV and CC )Split modeSplit / split lessSplit ratio1000 : 1Temperature programming7 stages / 8 platformsAuto samplerOptional
  • Mass Spectrometry Specifications
  • Model NoMS-89Mass range1.5 ~ 1000 amuIon – source temperature100 °C ~ 350 °CGC-MS interface temperature450 °CScan rateUpto 10000 amu/sStability± 0.10 amu / 48 hrs.Filament emission current0 ~ 350 µAEI source ionization energy5 eV ~ 250 eVResolutionUnit resolutionDetectorHigh energy dynode electron multiplierSensitivityFull scan ( S / N is ≥ 30 : 1 )VacuumTurbo molecular pump ( 67 L/s )

Portable Ion Chromatograph

Ion Chromatograph LPIC-A10

Ion Chromatograph LPIC-A10

  • Chromatographic Pump
  • TypeHigh-pressure and low –pulse double plunger tandem advection pumpMaximum Pressure35 MPa (PEEK)Flow Range0.001 to 9.999 mL/minFlow accuracy± 0.5 %Flow RepeatabilityRSD ≤ 0.1%Flow stability(0.2 to 0.5) mL/min, ≤ 3 % (0.5 to 1.0) mL/min, ≤ 2 % >1.0 mL/min, ≤ 2 %
  • Built-in Heat conduction type column heater
  • Temperature rangeRoom temperature +5°C to 60°C (41 to 140°F)Allowable deviation± 1°CTemperature stability≤ 0.5°C/h
  • Electromagnetic injection valve
  • Maximum pressure35 MPaControl modeBy stepper motorPower supply24 V (DC)