This model of HPLC gradient system is equipped with 2 sets of high-pressure constant-current pump and ultraviolet detector. Its microprocessor reciprocating plunger pump comes with functions such as high working pressure, low pulse rate. This gradient HPLC model meet the needs of conventional analysis to micro flow analysis.

High pressure pump Two sets of pump
Flow rate range 0.001 - 9.999 mL / min (0.001 mL / min increments)
Pressure range ≤ 0.5 MPa (0 - 42 MPa)
Flow rate precision RSD < 0.06 %
Pulsation ≤ 0.1 MPa (flow rate 1 ml / min, Pressure 5 - 10 MPa)
Qualitative analysis RSD 6 ≤ 0.1 % (naphthalene methyl alcohol solvent)
Quantitative repeatability RSD 6 ≤ 0.3 % (naphthalene methyl alcohol solvent)
Flow rate setting error Ss ≤ 2 % (1 mL / min water, 5 - 10 MPa room temperature)
Air tightness of the pump Pressure 42 MPa, 10 min time, Pressure drop < 0.5 MPa
Mixing accuracy ± 1 %
Mixing precision ± 0.2 %
Size (W × H × D) 450 × 300 × 160 mm
UV Detector <table class='table table-striped table-bordered table-responsive'><tr><td><strong>Wavelength range</strong></td><td>190 - 680 nm</td></tr><tr><td><strong>Flow cell</strong></td><td>8 µL</td></tr><tr><td><strong>Spectrum band width</strong></td><td>8 nm</td></tr><tr><td><strong>Baseline noise</strong></td><td>≤ ± 0.25 × 10<sup>-5</sup> AU (empty cell)</td></tr><tr><td><strong>Wavelength accuracy</strong></td><td>± 1 nm</td></tr><tr><td><strong>Baseline drift</strong></td><td>≤ ± 0.4 × 10<sup>-4</sup> AU (empty cell)</td></tr><tr><td><strong>Wavelength precision</strong></td><td>< 0.1 nm</td></tr><tr><td><strong>Limit of detection</strong></td><td>3 × 10<sup>-9</sup> g / mL (naphthalene methyl alcohol solvent)</td></tr><tr><td><strong>Size (W × H × D)</strong></td><td>450 × 300 × 160 mm</td></tr></table>
  • More precise optical system for the high precision and small offset
  • Improved light energy through the parallel double cone flow cell
  • Significant reduction in the static noise of the instrument and signal drift
This HPLC system has applications in the field of biochemistry, chemistry, pharmacology for chemical analysis, poison analysis and in protein food detection, medicine testing and polymer, environmental analysis.

Standard HPLC

HPLC system LHLC-B11

HPLC system LHLC-B11

  • High pressure pump
  • High pressure constant-current infusion pump
  • Infusion Way
  • Micro volume series dual-piston
  • Infusion pressure
  • 0 - 9999 Psi
HPLC system LHLC-B10

HPLC system LHLC-B10

  • High pressure pump
  • High pressure gradient pump
  • Peak operating pressure
  • 45 MPa, Stainless steel pump
  • Pressure
  • ± 0.02 MPa
HPLC system LHLC-A10

HPLC system LHLC-A10

  • High pressure pump
  • Single pump
  • Flow rate range
  • 0.001 - 9.999 mL / min (0.001 mL / min increment)
  • Pressure range
  • ≤ 0.5 MPa (0 - 42 MPa)

Analytical HPLC

Analytical HPLC System LAHS-B10

Analytical HPLC System LAHS-B10

  • High Pressure Constant Flow Pump
  • Double plunger reciprocating pump1 setFlow range0.001 to 10.00 mL/minSetting step size0.001 mL/minFlow accuracy≤ ± 0.3 %Flow stabilityRSD ≤ 00.1 %Maximum pressure42 MPaDisplay pressure error≤ 1.0 %Pressure fluctuation≤ 1.0 %Power consumption80 W
  • UV/Visible Detector
  • Wavelength range190 to 700 nmWavelength repeatability± 0.1 nmBand width8 nmResponse time0 to 4.9 secondsLinearity range1.8 AU (5 %)Baseline Noise≤ ± 1 × 10-5AUBaseline drift≤ 2.0 × 10-4AU/hMinimum detectable concentration4.0 × 10-9 g/mLDetector cellOptical path: 10 mm Volume: 10 µLLight sourceDeuterium lampPower consumption100 WVoltageAC 110 V/ 220 V ± 10 %, 50/60 HzPacking Dimensions (W × D × H)600 × 440 × 380 mm 2 × (600 × 410 × 300 mmGross weight54 kg
Analytical HPLC System LAHS-A10

Analytical HPLC System LAHS-A10

  • Pump specification
  • TypeQuaternary Low Pressure PumpFlow Rate Range0.001 to 10.000 ml/min, in 0.001ml incrementsFlow Accuracy≤ ± 0.2% (1.00 ml/min, H2 O, R.T., 8.5 MPa)Flow PrecisionRSD ≤ 0.06% (1.00 ml/min, H2 O, R.T., 8.5MPa)Maximum Pressure63 MPaPressure accuracy≤ ± 3.0 %Pressure pulsation≤ 0.1 MPaComposition rang0 to 100 %Composition precision≤ 0.3 % RSD
  • UV/VIS Detector
  • Wavelength Range190 to 800 nmWavelength Repeatability≤ 0.1 nmWavelength Accuracy≤ ± 1.0 nmBand Width6 nmResponse Time0.0 s to 5 sBaseline Noise≤ ± 0.25 × 10-5 AU(empty cell, 254 nm, 1.0s)Drift≤ 0.5 × 10-4 AU/h (empty cell, 254 nm, 60 min)Path length10 mm(analytical cell)Light SourceDeuterium Lamp and Tungsten Lamp
  • Data-processing Workstation
  • System contain the tables for regular samples Equipped with Internal and external standard calculation process Automatic display, print, export of data and starting of another program on completion on measurement Observe with the condition of 21 CFR part 11 directive of the FDA
Analytical HPLC-A7840

Analytical HPLC-A7840

  • Pump Specifications
  • Solvent deliveryDouble plunger series mode, floating plunger design, single stroke volume 30 µlPump Head10 ml stainless steel pump headFlow rate range0.001 to 9.999 ml / minFlow rate increment0.001 ml/minMaximum pressure42 MPaColumn size4.6 × 250 mm 5 µmPipeline Size1/16 inchFlow rate accuracy± 0.5 %Flow rate precisionRSD ≤ 0.1%Gradient accuracy %± 0.5 %Gradient repeatability %RSD ≤ 0.15%GLPAutomatic recording of pump sealing ring working timeDisplayLCD backlight displayPower80 WPower supply220 V ± 10%, 50 HzDimension398 × 149 × 267 mm
  • Detector Specifications
  • Wavelength range190 to 700 nmWavelength accuracy± 1 nmWavelength repeatability0.2 nmDeuterium lamp190 to 400 nmTungsten lamp400 to 700 nmBaseline noise ( Static ) Au1 × 10-5 AU/h at 254 nm with methanol/water= 80/20Baseline drift ( Static ) Au / h1.5 × 10-4 AU/h at 254 nm with methanol/ water= 80/20Time constants16 grades adjustableDisplayLCD backlight displayGLPTotal number of lights, lamp replacement time and product serial numberPower75 WPower supply220 V ± 10%, 50 HzDimension (L × W × H)398 × 149 × 267 mm
Analytical HPLC-A7835

Analytical HPLC-A7835

  • Pump Specifications
  • Model NoHPP-B1Flow rate range0.0001 to 10 ml / minMaximum pressure42 MPa , 0 to 5 ml / minColumn size4.6 × 250 mmSample size5 μmPressure pulsation≤ 1.5 % at 10 MPaFlow rate accuracy± 0.2 % at 1 ml / minFlow rate repeatability %≤ 0.15Gradient accuracy %± 2Gradient repeatability %≤ 0.2Dimension450 × 260 × 190 mmPower supply90 V to 240 VWeight11.2 kg
  • Detector Specifications
  • Model NoHPD-B2Wavelength range190 to 800 nmWavelength accuracy± 0.75 nmWavelength repeatability0.2 nmBand width8 nmOptical path10 nmDeuterium lamp190 to 400 nmTungsten lamp400 to 800 nmBaseline noise ( Static ) Au≤ ± 0.5 × 10 -5Baseline drift ( Static ) Au / h≤ 1.5 x 10 -4Minimum detection limit≤ 4 x 10 -8Power100 WPower supply90 V / 50 to 60 HzDimension460 × 260 × 190 mmWeight9.2 kg